Instrument number:2019A437 | Specification:PHOTON 100 CMOS 2D detector; Mo, Cu double target; Helium refrigeration system
| Manufacturer: Bruck, Germany |
Model:D8 Venture | Country of Manufacture: Germany | Date of purchase: 2020/02/24 |
Place: F-109,State Key Laboratory of Fine Chemicals |
|
|
Main specifications and technical indicators: Single crystal diffraction system, one or two X-ray sources; Dimensions: 202 x168 x129cm(H x W x D); Pixel size (μm) : 96 reduction times: 1:1(no taper); Quantum gain (electron, Mo Kα) : 450; SNR: 1.8(t<10s), 1.7(t =120s) X-ray source: Sealed tube; Available wavelength: copper, molybdenum; Cooling water supply: zei small 4 L /min(10°C to 20°C) Power supply: single-phase 208V to 240 V or three-phase 120 V, 230 V, 240 V; 47Hz to 63Hz. |
Main functions and features: A nondestructive analytical test technique that provides the internal structure of a single crystalline substance, including cell size, bond length, bond Angle, and spatial arrangement of atoms.
|
Main accessories and configuration: Single crystal diffraction system, one or two X-ray sources; Detection system detector: PHOTON100; X-ray source: Sealed tube; IμS microfocus source available wavelength: copper, molybdenum, silve.